Technical Alarms (INOPs)
Alarm Text
ECG/ARRH ALARM(S)
OFF
Source - MX40
ECG LEADS OFF
Note This INOP may also
be configured to display as
a Red or Yellow Technical
Alarm.
Source - MX40
<electrode> LEAD OFF
Source - MX40
LEADSET UNPLUGGED
Source - MX40
LEADSET LIFE
LOCAL AUDIO OFF
Source - MX40
Note —
This is normal
operation in Telemetry
Mode.
NBP INTERRUPTED
Source - Cableless
Measurement Device
NBP MEASURE FAILED
Source - Cableless
Measurement Device
5-16
Alarms
Priority
Condition
Soft
ECG is turned off.
Red or
Multiple leads are off.
Yellow or
Hard
Technical
Alarm
Hard
Single lead is off.
If primary lead is MCL,
lead will be identified as
V/C in INOP text.
Hard
Patient cable has been
unplugged from the
MX40.
Incompatible leadset
attached to patient cable.
Soft
The single-patient use
leadset has exceeded its
limit of 25 cycles.
Soft
There is no alarm audio
notification when operating
in Telemetry Mode.
Hard
The preset maximum time
for the total measurement
has been exceeded.
Hard
Measurement values
cannot be derived.
What to do
Turn on ECG.
Re-attach ECG leads to patient..
Re-attach ECG leads to patient.
Re-attach the patient cable.
Replace the leadset.
Replace with new leadset.
Change to Monitor Mode.
Reduce patient movement and
avoid interaction with the cuff
and tubing.
Attach cuff to new location on
patient.
Replace cuff.