Alarm Text
CHECK PAIRING
Source - MX40
cl NBP Batt Low
Source - Cableless
Measurement Device
cl NBP Batt Empty
Source - Cableless
Measurement Device
cl NBP DISCONNECT
Source - Cableless
Measurement Device
cl SpO
Batt Low
2
Source - Cableless
Measurement Device
cl SpO
Batt Empty
2
Source - Cableless
Measurement Device
cl SpO
DISCONNECT
2
Source - Cableless
Measurement Device
!!!/!! CUFF NOT
DEFLATED
Source - Cableless
Measurement Device
!!!/!! CUFF OVERPRESS
Source - Cableless
Measurement Device
Priority
Condition
Yellow
There is a problem with
Technical
device pairing.
Alarm
When the MX40 is
wirelessly paired with an
X2 patient monitor (no
label) docked with a
larger networked MP
series monitor, and the
network connection is
lost.
Hard
CL NBP Pod weak battery
condition.
Severe
CL NBP Pod empty battery
condition. Monitoring is not
possible.
Hard
CL NBP Pod is not
connected with the MX40.
Hard
CL SpO
Pod weak battery
2
condition.
Severe
CL SpO
Pod empty
2
battery condition.
Monitoring is not possible.
Hard
CL SpO
Pod is not
2
connected with the MX40.
Severe
Cuff pressure has
exceeded the specified
safety limit.
Severe
Cuff pressure has
increased above
overpressure safety limits.
Technical Alarms (INOPs)
What to do
Check that the bedside
monitor or cableless
measurement device is
correctly paired.
Select the correct device to be
paired.
Charge CL NBP Pod.
Replace CL NBP Pod.
Recharge depleted CL NBP
Pod.
Resolve interference
condition.
Reduce range between CL
NBP Pod and MX40.
Charge CL SpO
Pod.
2
Replace CL SpO
Pod.
2
Recharge depleted SpO
Resolve interference
condition.
Reduce range between CL
SpO
Pod and MX40.
2
Remove cuff and tubing and
expel air.
Remove cuff and tubing and
expel air.
Alarms
Pod.
2
5-15