Technical Alarms (INOPs)

Technical Alarms (INOPs)
Alarm Text
BATTERY LOW T
Source - MX40
CANNOT ANALYZE ECG
Source - MX40 (IIC iX
only) and Information
Center
5-14
Alarms
Technical Alarms, or INOPs (inoperative conditions), are sourced at the
MX40, the ST/AR algorithm running at the Information Center, or the
IntelliVue Patient Monitor. They identify inoperative conditions (that is
conditions where the system is not operating properly and therefore cannot
measure or detect alarm conditions reliably). There are four levels of
Technical Alarms:
Severe - Monitoring and alarm generation are disabled. Visual alarm
indicator on the MX40. Audible tone at the Information Center. Must be
acknowledged by a clinician.
Hard - Monitoring and alarm generation are disabled. Visual alarm
indicator on the MX40. Audible tone at the Information Center.
If the hard INOP is "latched", the sound will be silenced, but the
message will remain on the display until resolution of the offending
condition.
Soft - Monitoring and alarms remain active. Visual alarm indicator on
the MX40 and at the Information Center. No audible tones are
generated at the Information Center
Red/Yellow - Replace Battery and ECG Leads Off INOPs may be
configured to display as either Red or Yellow Technical Alarms.
Note - The ECG Leads Off INOP will initially display as a cyan
technical alarm until a valid ECG signal is obtained.
In the following table, technical alarms are listed alphabetically.
Priority
Soft
Hard
Condition
There is less than 15
minutes of monitoring
time remaining (AA
batteries).
Lithium-ion battery level
is < 10% or has <30
minutes remaining time.
Arrhythmia algorithm
cannot reliably analyze the
ECG data on any
monitored leads.
What to do
Replace batteries promptly to
avoid shutdown and cessation
of monitoring.
Insert a charged lithium-ion
battery pack.
Assess the lead selections,
initiate relearn, and validate
analyzed rhythm. Check other
INOPs for possible source of
problem.