Physiologic Alarms
Alarm Text
*/** HR High
*/** HR Low
*/** IRREGULAR HR
* MISSED BEAT
* MULTIFORM PVCs
* NON-SUSTAIN VT
*Nurse Call
* PACER NOT CAPTURE
* PACER NOT PACING
* PAIR PVCs
* PAUSE
* PVCs /MIN HIGH
5-12
Alarms
Priority
Condition
Yellow
Heart Rate (yyy) greater than the upper
Heart rate limit (xxx).
Yellow
Heart Rate (yyy) lower than the lower
Heart Rate limit (xxx).
Yellow
Consistently irregular rhythm (irregular
R-R intervals).
Yellow
No beat detected for 1.75 x average R-R
interval for Heart Rate greater than 120, or
no beat for 1 second with Heart Rate
greater than 120 (non-paced patient only).
Yellow
The occurrence of two differently shaped
Vs, each occurring at least twice within the
last 300 beats as well as each occurring at
least once within the last 60 beats.
Yellow
A run of Vs having a ventricular Heart
Rate greater than V-Tach limit but lasting
for less than the V-Tach Run limit.
Yellow
The patient has pressed the
Multi-Function Button on the MX40.
Yellow
No QRS for 1.75 x the average R-R
interval with Pace Pulse (paced patient
only).
Yellow
No QRS and Pace Pulse for 1.75 x the
average R-R interval (paced patient only).
Yellow
Two consecutive PVCs between
non-PVCs.
Yellow
No QRS detected for x seconds. Choices
of >1.5 to 2.5 seconds.
Yellow
PVCs within one minute exceed by the
PVCs/min limit (xxx).
Source
ST/AR Basic
& Enhanced
Arrhythmia
ST/AR Basic
& Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR Basic
& Enhanced
Arrhythmia
ST/AR Basic
& Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR Basic
& Enhanced
Arrhythmia