Download  Print this page

Advertisement

Alarm Text
***EXTREME TACHY
*** VENT FIB/TACH
*** V-TACH
** dQTc
** MULTI ST
** NBP High
** NBP Low
** QTc High
** RR HIGH
Note —
Only available with
IIC iX.
** RR LOW
Note —
Only available with
IIC iX.
**SpO
T HIGH
2
** SpO
T LOW
2
** ST-LEAD HIGH
** ST-LEAD LOW
** ST ELEV
*/**AFIB
Priority
Condition
Red
Extreme Tachycardia.
Heart Rate (yyy) greater than Extreme
Tachy limit
Red
Ventricular Fibrillation.
Fibrillatory wave (sinusoidal wave
between 2-10 Hz) for 4 consecutive
seconds
Red
Ventricular Tachycardia.
Consecutive PVCs greater than or equal
to V-Tach Run limit and Heart Rate
greater than V-Tach limit (xxx)
Yellow
dQTC value has exceeded the dQTC high
limit for > 5 minutes for 2 consecutive time
periods.
Yellow
Two contiguous ST leads have values
exceeding elevation or depression alarm
limits for > 60 sec.
Yellow
High limit has been exceeded for high
pressure limit.
Yellow
Low limit has been exceeded for low
pressure limit.
Yellow
QTc value has exceeded the QTc high
limit for > 5 minute for2 consecutive time
periods.
Yellow
The respiration rate has exceeded the
high alarm limit.
Yellow
The respiration rate has dropped below
the low alarm limit.
Yellow
High SpO
.
2
SpO
value (yyy) greater than high SpO
2
limit (xxx)
Yellow
Low SpO
.
2
SpO
value (yyy) less than low SpO
2
(xxx).
Yellow
ST segment is elevated.
Yellow
ST segment is depressed.
Yellow
Two contiguous leads have ST values
exceeding the STEMI limits.
Yellow
Atrial fibrillation waveform detected.
Physiologic Alarms
Source
ST/AR Basic
& Enhanced
Arrhythmia
ST/AR Basic
& Enhanced
Arrhythmia
ST/AR Basic
& Enhanced
Arrhythmia
ECG/QT
ECG/ST
NBP
NBP
ECG/QT
RESP
RESP
SpO
2
2
SpO
2
limit
2
ECG/ST
ECG/ST
ECG/ST
ST/AR
Enhanced
Arrhythmia
Alarms