AA Battery Charge Status

Approximate
Approximate
Battery
Time
Life
Remaining
Remaining
(ECG only)
100%
~ 24 hours
75%
< 18 hours
50%
< 12 hours
25%
< 6 hours
10%
< 2 hours
Low battery level
< 30 minutes
to
replace/charge
battery level
Replace/charge
< 10 minutes
battery level
Approximate
Functionality
Time
Disabled
Remaining
(ECG & Spo2
Continuous)
~ 9 hours
None
< 7 hours
None
< 5 hours
None
< 2 hours
None
< 1 hours
None
< 30 minutes
SpO
and
2
short-range radio
are disabled.
Display is at half
brightness.
< 10 minutes
Device shutdown 1 Red
Battery Information
Battery
Indicator
LCD
Segments
5 Green
4 Green
3 Green
2 Green
1 Green
1 Red
Red Battery
Icon
Audio
Red Battery
Icon