Battery Information

Battery Charge Status

4-20
Basic Operation
Disposal of Batteries
When disposing of batteries, follow local laws for proper disposal. Dispose
of batteries in approved containers. If local regulations require you to
recycle batteries, recycle batteries in accordance with those regulations.
The battery charge indicator displays in the Status Area and communicates
the remaining battery charge time when using both AA batteries or the
rechargeablelithium-ion battery.
When the MX40 is initially powered-on, it takes approximately 25 seconds
for the indicator to populate. During this time, the indicator displays a ? in
the battery icon.
In order to guarantee overall device performance, certain functionality is
disabled when the battery charge reaches critical levels. See the tables
below for additional information about battery status.

Lithium-ion Rechargeable Battery Charge Status

Approximate
Approximate
Battery
Time
Life
Remaining
Remaining
(ECG only)
100%
~ 25 hours
75%
< 19 hours
50%
< 13 hours
25%
< 6 hours
10%
< 3 hours
Low battery level
< 30 minutes
to replace/charge
battery level
Replace/charge
< 10 minutes
battery level
Approximate
Functionality
Time
Disabled
Remaining
(ECG & Spo2
Continuous)
~ 14 hours
None
< 10.5 hours
None
< 7 hours
None
< 3.5 hours
None
< 1.5 hours
None
< 30 minutes
SpO
and
2
short-range radio
are disabled.
Display is at half
brightness
< 10 minutes
Device shutdown 1 Red
Battery
Indicator
LCD
Segments
5 Green
4 Green
3 Green
2 Green
1 Green
1 Red
Red Battery
Icon
Audio
Red Battery
Icon