Understanding Settings

Understanding Settings

Changing Measurement Settings

ECG Settings at the MX40

Setting
Alarm Limits
Primary
(used for arrhythmia analysis only)
(Set at IIC Release N or IIC iX. View only.)
Secondary
(used for arrhythmia analysis only)
(Set at IIC Release N or IIC iX. View only.)
Paced Mode
(Set at IIC Release N or IIC iX)
Adjust Size
Arrhythmia
Lead Placement
4-12
Basic Operation
Each aspect of how the MX40 works and looks is defined by a setting. There
are a number of different categories of settings, including:
Screen Settings - to define the selection and appearance of elements on
each individual display screen.
Measurement Settings - to define setting unique to each measurement,
e.g. high and low alarm limits.
Monitor Settings -including settings that affect more than one
measurement or display screen, for example alarm volume and alarm
pause time.
You must be aware that, although many settings can be changed during
use, permanent changes to settings can only be done in Configuration
Mode. All settings are restored to their default setting when the patient is
discharged or the MX40 is powered off.
Each measurement has a setup menu in which you can adjust its settings.
You enter the setup menu by selecting the measurement numeric.
Description
Heart Rate alarm limits can be viewed locally at
the MX40. Limits set at the Information Center
(Release N or later or iX) are reflected at the
MX40 when connected on the network.
I, II, III, aVR, aVL, aVF, V1-V9, MCL, V3R, V4R,
V5R. Available waveforms are based on lead set
type. Lead II is the default.
I, II, III, aVR, aVL, aVF, V1-V9, MCL, V3R, V4R,
V5R. Available waveforms are based on lead set
type. Lead V is the default.
On, Off
Set ECG gain to x1/2, x1, x2, x4
Initiate an Arrhythmia Relearn; View Arrhythmia
Alarm Limits; Turn Arrhythmia Annotation On/Off.
Set EASI, Standard