Measurement Specifications
Characteristic
Time to Alarm for Tachycardia
Pacemaker Pulse Rejection Capability
Range and Accuracy of Heart Rate Meter
Alarm Limit Range
Resolution of Alarm Limit Settings
Alarm Limit Accuracy
Time to Alarm for Cardiac Standstill
Time to Alarm for Low Heart Rate
Time to Alarm for High Heart Rate
Alarm Silencing
ECG Waveform Display Time Base
Accuracy
Channel Width
Trace Width
Aspect Ratio
Input Signal Reproduction Accuracy:
Overall Error
Frequency Response: Sinusoidal
15-28 Safety Standards & Specifications
Performance Disclosure/Specification (in
italics)
The ranges of time to alarm using test waveforms as
indicated in ANSI/AAMI EC13 Sec. 4 1. 2. 1 (g) are 4 to 5
seconds.
Rejects pace pulses using test waveforms as indicated in
ANSI/AAMI EC13 Sec. 4. 1. 4.1 (with amplitude from +/- 2
to +/- 700 mV, width from 0.1 to 2.0 ms).
Meets the ANSI/AAMI EC13 Section 4.2.7 recommended
minimum range and accuracy.
Heart rate range is 15 - 300 b/min for adults patients and
15-350 b/min for pediatric patients with accuracy of
the range. (Note: for rates equal to or less than 15, the
displayed heart rate is 0).
Meets the ANSI/AAMI EC13 Section 4.2.8.1 standard.
Lower alarm limit is 15 -295. Upper alarm limit is 20 - 300.
Meets the ANSI/AAMI EC13 Section 4.2.8.2 standard. The
resolution is
5 b/min.
Meets the ANSI/AAMI EC13 Section 4.2.8.3 standard.
Error less than ± 10% or ± 5b/min
Meets the ANSI/AAMI EC13 Section 4.2.8.4 standard:
maximum alarm time <10 seconds, using the test
waveforms as indicated.
Meets the ANSI/AAMI EC13 Section 4.2.8.5 standard:
maximum alarm time <10 seconds, using the test
waveforms as indicated.
Meets the ANSI/AAMI EC13 Section 4.2.8.6 standard:
maximum alarm time <10 seconds, using the test
waveforms as indicated.
The time required for reactivation of a latched, silenced
alarm is 3 minutes
Meets the ANSI/AAMI EC13 Section 4.2.9.6 standard:
maximum error = +/-10%.
Meets the ANSI/AAMI EC13 Section 4.2.9.7(a) standard:
minimum = 30mm.
Meets the ANSI/AAMI EC13 Section 4.2.9.7(b) standard:
maximum = 1.0mm.
Meets the ANSI/AAMI EC13 Section 4.2.9.7(f) standard: 0.4
+ 0.08 s/mV.
Meets the ANSI/AAMI EC13 Section 4.2.9.8(a) standard:
maximum = +/- 20%.
Meets the ANSI/AAMI EC13 Section 4.2.9.8(b) standard:
0.67 to 40 Hz
(3 db down).