Download  Print this page

Ecg Performance Disclosure/specifications - Philips IntelliVue MX40 Instructions For Use Manual

Hide thumbs

Advertisement

Parameter
Pacer Rejection
Performance
(Pace pulses with no
tails).
EMC Performance Limits,
radiated immunity
ECG Patient Cable
Disconnection Safety

ECG Performance Disclosure/Specifications

Characteristic
Heart Rate Averaging Method
Heart Rate Meter Accuracy and Response
to Irregular Rhythm
Response Time of Heart Rate Meter to
Change in Heart Rate
Specification
Positive pacers
1
Amplitude
+2 to +700 mV
+2 to +500 mV
+2 to +400 mV
Negative pacers
1
Amplitude
-2 to -700 mV
-2 to -500 mV
-2 to -400 mV
1
Philips does not claim, verify, or validate support
for all available pacemakers.
Meets Essential Performance.
All ECG connections are patient safe within 750
msec of patient cable removal, with patient leakage
current <10
Exception: Leadset detection pins
are protected mechanically to prevent patient
contact.
Performance Disclosure/Specification (in
italics)
Two different methods are used:
Normally, heart rate is computed by averaging the 12
most recent RR intervals.
If each of 3 consecutive RR intervals are greater than
1200 milliseconds (i.e. rate less than 50 b/min) for adult
and pediatric patients, then the 4 most recent RR
intervals are averaged to compute the HR.
Provides correct heart rates (80, 60, 120, 90 b/min) using
test waveforms as indicated in ANSI/AAMI EC13
Sec. 4. 1. 2. 1 (e).
For a rate increase, the average time to reach the specified
heart rate using test waveforms as indicated in ANSI/AAMI
EC13 Sec. 4. 1. 2. 1 (f) is 10 seconds. For a rate drop, the
average time is 7 seconds.
Safety Standards & Specifications
Measurement Specifications
Width
0.1, 0.2, 0.5 and 1.0 ms
1.5 ms
2 ms
Width
0.1, 0.2, 0.5 and 1.0 ms
1.5 ms
2 ms
15-27

Advertisement

Table of Contents

   Also See for Philips IntelliVue MX40

Comments to this Manuals

Symbols: 0
Latest comments: