Measurement Specifications

Measurement Specifications

ECG

15-26 Safety Standards & Specifications
Parameter
ECG channel transmitted
Leads
3 electrodes
5 electrodes
5 electrodes, EASI
6 electrodes
Resolution
ECG Input
Input Impedance
Input Dynamic Range
DC Offset Range
CMRR
Bandwidth +/- 3 dB
Gain Accuracy
Noise Referred to ECG
Input (Peak-to-Peak)
Lead Wires
Time to baseline recovery
from Defibrillator
Specification
Channel #1 = I, II, or III
Channel #1 = II
Channel #2 = III
Channel #3 = MCL
Channel #1 = Va-i
Channel #2 = Va-s
Channel #3 = Ve-s
Channel #1= II
Channel #2 = III
Channel #3 = MCLa
Channel #4 = MCLb
5 V
Differential, defibrillator protected against 360
joules discharge into a 100 ohm load
> 5 megohms (@ 10 Hz
+/- 9 mV
+/- 320 mV
> 90 dB @ 50, 60 Hz
0.05 to 40 Hz
+/- 5% at 25
o
C (77
o
F)
AAMI: 30 V (as per AAMI EC 13)
3, 5 or 6-wire patient cable compatible with
IntelliVue Patient Monitor, AAMI/IEC color codes
AAMI: 5 s max (until ECG wave is on display but
not yet centered, monitoring bandwidth)