Download  Print this page

Electromagnetic Emissions; Electromagnetic Immunity - Philips IntelliVue MX40 Instructions For Use Manual

Hide thumbs

Advertisement

Electromagnetic Compatibility

Electromagnetic Emissions

Electromagnetic Immunity

Immunity Test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
15-10 Safety Standards & Specifications
Warning
Using accessories other than those specified may result in increased
electromagnetic emission or decreased electromagnetic immunity of the
monitoring equipment.
Emissions Test
Radio Frequency (RF)
emissions
RF emissions CISPR
11
Harmonized emissions Not Applicable
Voltage
fluctuations/Flicker
emissions
IEC 61000-3-3
The MX40 is suitable for use in the specified electromagnetic environment.
The user must ensure that it is used in the appropriate environment as
described below.
IEC 60601-1-2 Test
Level
±6 kV contact
±8 kV air
3 A/m
Compliance
Avoiding Electromagnetic
Interference
Group 1
TheMX40 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
Class B
The MX40 is suitable for use in all
establishments.
Device is battery powered only
Not Applicable
Compliance Level
±6 kV contact
±8 kV air
3 A/m
Electromagnetic
Environment Guidance
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Power frequency magnetic fields
should be a t levels characteristic
of a typical location in a typical
commercial and/or hospital
environment

Advertisement