Unit Settings - NBP Setup
Control
Function
Set patient type used for NBP
Patient
alarm limits.
Type
Set NBP alarm notification.
NBP Alarm
Increment/decrement NBP high
Systolic
alarm limit by 5.
High
Increment/decrement NBP low
Systolic
alarm limit by 5.
Low
Increment/decrement NBP high
Diastolic
alarm limit by5.
High
Increment/decrement NBP low
Diastolic
alarm limit by 5.
Low
Increment/decrement NBP high
Mean High
alarm limit by 5.
Increment/decrement NBP low
Mean Low
alarm limit by 5.
Settings
Adult
Pediatric
Systolic or Diastolic
Systolic
Diastolic
Mean
Off
Limit Maximum is 260
Limit Minimum is 10
Limit Maximum is 260
Limit Minimum is 10
Limit Maximum is 260
Limit Minimum is 10
Limit Maximum is 260
Limit Minimum is 10
Limit Maximum is 260
Limit Minimum is 10
Limit Maximum is 260
Limit Minimum is 10
Monitoring with the MX40 at the Information Center
Unit Configurable Settings
Factory
Default
Adult
Systolic or
Diastolic
160 Adult
120 Pediatric
90 Adult
70 Pediatric
90 Adult
70 Pediatric
50 Adult
40 Pediatric
110 Adult
90 Pediatric
60 Adult
50 Pediatric
11-19