Unit Settings - Telemetry Setup
Control
Function
Enable/disable the
Pleth Wave
transmission of the Pleth
wave and its subsequent
display to the Information
Center. For Continuous
mode only.
Enable/disable the
Pulse
transmission of the Pulse
rate and its subsequent
display to the Information
Center. For Continuous
mode only.
Turn SpO
SpO
Alarm
2
at the Information Center.
Increment/decrement
SpO
Limits
2
SpO
High
(in %).
Increment/decrement
SpO
Limits
2
SpO
Low
(in %).
Settings
enable
disable
Note —
during 6-lead monitoring, the
Pleth wave will replace the Vb
wave in the Patient Window.
enable
disable
alarms on/off
enable (on)
2
disable (off)
Limit maximum is 100. Limit
high alarm limit by 1
minimum is 51 (adult) or 31
2
(pediatric). High and low limit
must be at least 1% apart.
Limit maximum is 99. Limit
low alarm limit by 1
minimum is 50 (adult) or 30
2
(pediatric). High and low limit
must be at least 1% apart.
Monitoring with the MX40 at the Information Center
Unit Configurable Settings
Factory
Default
disable (Pleth
wave is not
displayed.)
When enabled,
disable (Pulse
rate is not
displayed.)
enable
100 (adult,
pediatric)
90 (adult,
pediatric)
11-17