Unit Configurable Settings
Unit Settings - Telemetry Setup
Control
Patient Type
Telemetry/
Multi-Function
Button
Standby
Duration
Enable Remote
Suspend
Suspend
Duration
Battery Gauge
on Information
Center
11-14 Monitoring with the MX40 at the Information Center
Function
Set patient type used for
SpO
alarm limits.
2
Determine the Information
Center response when
Telemetry Button is
pressed.
Sets the standby duration
on the MX40 when
Standby is initiated from
the Information Center.
Note —
When Standby
is initiated from the MX40,
the duration is always
Infinite.
Enable/disable alarm
pause/suspend at the
MX40.
Sets the alarm suspend
duration time for each
assigned device on the
Information Center.
Display/disable a battery
gauge for each assigned
device on the Information
Center.
Settings
Adult
Pediatric
Nurse Call - generate nurse
call alarm that can be
retrieved from Alarm Review
for later use.
Record - generate a recording
strip
Both - generate nurse call
alarm and recording strip
None
Infinite
10 minutes
20 minutes
30 minutes
1 hour
2 hours
3 hours
4 hours
enable
disable
1, 2, or 3 minutes
enable
disable
Factory
Default
Adult
Nurse Call
Infinite
disable
2 minutes
enable (battery
gauge is
displayed)
Note —
the
battery gauge is
always displayed
on the MX40.