Patient Configurable Settings in Telemetry Setup
Patient-Configurable Settings in Telemetry Setup
Control
Suppress
SpO
INOPs
2
with NBP
Pleth Wave
Pulse
SpO
Alarm
2
Unit Settings
11-12 Monitoring with the MX40 at the Information Center
Function
Setting Choices
Enable/disable the
enable
SpO
algorithm to
2
disable
suppress sending
technical alarms from
the MX40 during an
NBP measurement for
60 seconds.
Warning
If you measure SpO
2
on a limb that has an
inflated NBP cuff, a
non-pulsatile SpO
2
technical alarm can
occur. If the monitor is
configured to suppress
this alarm, there can be
a delay of up to 60
seconds in indicating
critical patient status,
such as sudden pulse
loss or hypoxia.
Enable/disable the
enable
transmission of the
disable
Pleth wave (and its
Note —
subsequent display) to
the Information Center.
wave replaces the Vb wave in the
For Continuous SpO
Patient Window during 6-lead
2
mode only.
monitoring.
Enable/disable display
enable
of the Pulse rate at the
disable
Information Center. For
Continuous SpO
mode
2
only.
enable (on)
Turn SpO
alarms
2
on/off at the Information
disable (off)
Center and the MX40.
Change current settings
(none)
back to last saved
clinical unit settings.
When enabled, the Pleth
Factory
Default
enable
disable (Pleth is
not displayed.)
disable (Pulse rate
is not displayed.)
enable