Patient-Configurable Settings in Telemetry Setup
Control
Function
Determines the
Telemetry/
Information Center
Multi-Function
response when the
Button
Multi-Function Button is
pressed.
Sets the appearance of
Fixed Pacer
the pacer spikes to a
Amplitude
fixed size as they
appear in the patient
window.
Enable/disable the
SpO
Enabled
2
SpO
2
the Information Center
and the MX40.
Determine the MX40
SpO
Mode
2
SpO
2
behavior.
Note —
and Pleth Wave are not
available in Spot Check
(Manual) mode.
Setting Choices
Nurse Call - generate nurse call
alarm that can be retrieved from
Alarm Review for later use.
Record - generate a recording strip
Nurse Call and Record - generate
nurse call alarm and recording strip
None
enable
disable
Note —
appearance of the pacer spikes on
the MX40.
enable
measurement at
disable
Spot Check (Manual)- Provides
measurement
manual measurements so the
clinician can check as needed.
Measurement can be initiated at the
Pulse Rate
MX40 from the SmartKeys menu or
the SpO
Information Center by selecting the
Spot Check SpO
Window.
Continuous - Sends an SpO
parameter value to the Information
Center every second. If selected,
Pulse Rate and Pleth Wave may also
be sent.
Monitoring with the MX40 at the Information Center
Patient Configurable Settings in Telemetry Setup
this does not affect the
Setup menu and at the
2
icon in the Patient
2
2
Factory
Default
Nurse Call
disable
enable
Spot Check
(Manual)
11-11