Monitoring during Leads Off
6-26
ECG and Arrhythmia Monitoring
Warning
Since Relearn happens automatically, if learning takes place during
ventricular rhythm, the ectopics can be incorrectly learned as the normal
QRS complex. This can result in missed detection of subsequent events of
V-Tach and V-Fib. For this reason, you should:
1
Respond promptly to any technical alarm.
2
Ensure that the arrhythmia algorithm is labeling beats correctly.