Download  Print this page
   
1
2
Table of Contents
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
185
186
187
188
189
190
191
192
193
194
195
196
197
198
199
200
201
202
203
204
205
206
207
208
209
210
211
212
213
214
215
216
217
218
219
220
221
222
223
224
225
226
227
228
229
230
231
232
233
234
235
236
237
238
239
240
241
242
243
244
245
246
247
248
249
250
251
252
253
254
255
256
257
258
259
260
261
262
263
264
265
266
267
268
269
270
271
272
273
274
275
276
277
278
279
280
281
282
283
284
285
286
287
288
289
290
291
292
293
294
295
296
297
298
299
300
301
302
303
304
305
306
307
308
309
310
311
312
313
314
315
316
317
318
319
320
321
322
323
324
325
326
327
328
329
330
331
332
333
334
335
336
337
338
339
340
341
342
343
344
345
346
347
348
349
350
351
352
353
354
355
356
357
358
359
360
361
362
363
364
365
366
367
368
369
370
371
372
373
374
375
376
377
378
379
380
381
382
383
384
385
386
387
388
389
390
391
392
393
394
395
396
397
398
399
400
401
402
403
404
405
406
407
408
409
410
411
412
413
414
415
416
417
418
419
420
421
422
423
424
425
426
427
428
429
430
431
432
433
434
435
436
437
438
439
440
441
442
443
444
445
446
447
448
449
450
451
452
453
454
455
456
457
458
459
460

Advertisement

INOP Message, Indication
CO₂ NO TUBING
Numeric is replaced by -?-
INOP tone
CO₂ OVERRANGE
Numeric is replaced by -?-
INOP tone
CO₂ PUMP OFF
Numeric is replaced by -?-.
CO₂ PURGING
Numeric is replaced by -?-
INOP tone
CO₂ SENS.WARMUP
Numeric is replaced by -?-
Microstream CO
: INOP tone
2
Mainstream CO
: no INOP tone
2
CO₂ UPDATE FW
Numeric is replaced by -?-
INOP tone
CO₂ WAIT CAL2
Numeric is replaced by -?-
CO₂ ZERO FAILED
Numeric is replaced by -?-
INOP tone
CO₂ ZERO REQU'D
Numeric is replaced by -?-
INOP tone
CO₂CHANGE SCALE
CO₂ZERO RUNNING
SO2 INOPs
INOP Message, Indication
<SO₂ Label> CAL FAILED
Numeric is replaced by -?-
INOP tone
<SO₂ Label> CAL MODE
Numeric is replaced by -?-
INOP tone
<SO₂ Label> CANNOT
MEAS
Numeric is replaced by -?-
INOP tone
<SO₂ Label> CONN
OPTMOD
Numeric is replaced by -?-
INOP tone
What to do
Either the sample line is disconnected, or an incorrect line is attached. Check the connection. If
necessary, connect another sample line (Use only the approved accessories).
If you silence this INOP, the measurement will be switched off.
The CO
value is higher than the measurement range. If you suspect a false high value, contact your
2
service personnel.
The pump has been switched off for fifteen minutes. To switch it on again, select Pump On in the
Setup CO₂ menu.
The Filterline is being purged to remove an occlusion in the line or airway adapter. If the occlusion is
removed, the INOP will disappear. If not, the INOP CO₂ OCCLUSION is displayed.
Wait until the sensor reaches operating temperature and the INOP disappears.
The software in the Measurement Extension does not match the software in the MMS. Contact your
service personnel.
Calibration on the first calstick cell is complete. Place the transducer on the other calstick cell and start
the CAL2 calibration cycle.
An error occurred during the last zero calibration. Check the airway adapter and clean, if necessary.
Perform another zero calibration. If the INOP persists, contact your service personnel.
Perform zero calibration for the CO
The CO
wave is clipped. Select a more appropriate wave scale to display the whole wave.
2
Wait until zero calibration is finished.
What to do
The calibration failed. Check the catheter-to-Optical-Module connection. Manually restart the
calibration. Try another catheter and Optical Module. If the catheter is already inserted, perform an in-
vivo calibration.
Pre-insertion calibration is complete, but the catheter tip is still inside the optical reference. The
catheter is now ready for insertion.
The signal is out of the normal range, and no oxygen saturation can be derived. Perform an in-vivo
calibration. If the INOP persists, try another Optical Module and catheter.
The Optical Module was disconnected during data storage. Reconnect the Optical Module for at least
20 seconds.
3 Patient Alarms and INOPs
sensor. If the INOP persists, contact your service personnel.
2
81

Advertisement

Table of Contents

Comments to this Manuals

Symbols: 0
Latest comments: