Download  Print this page

Inspecting The Cables And Cords; Maintenance Task And Test Schedule - Philips IntelliVue MX800 Instructions For Use Manual

Patient monitoring.
Hide thumbs
   
1
2
Table of Contents
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
185
186
187
188
189
190
191
192
193
194
195
196
197
198
199
200
201
202
203
204
205
206
207
208
209
210
211
212
213
214
215
216
217
218
219
220
221
222
223
224
225
226
227
228
229
230
231
232
233
234
235
236
237
238
239
240
241
242
243
244
245
246
247
248
249
250
251
252
253
254
255
256
257
258
259
260
261
262
263
264
265
266
267
268
269
270
271
272
273
274
275
276
277
278
279
280
281
282
283
284
285
286
287
288
289
290
291
292
293
294
295
296
297
298
299
300
301
302
303
304
305
306
307
308
309
310
311
312
313
314
315
316
317
318
319
320
321
322
323
324
325
326
327
328
329
330
331
332
333
334
335
336
337
338
339
340
341
342
343
344
345
346
347
348
349
350
351
352
353
354
355
356
357
358
359
360
361
362
363
364
365
366
367
368
369
370
371
372
373
374
375
376
377
378
379
380
381
382
383
384
385
386
387
388
389
390
391
392
393
394
395
396
397
398
399
400
401
402
403
404
405
406
407
408
409
410
411
412
413
414
415
416
417
418
419
420
421
422
423
424
425
426
427
428
429
430
431
432
433
434
435
436
437
438
439
440
441
442
443
444
445
446
447
448
449
450
451
452
453
454
455
456
457
458
459
460

Advertisement

34 Maintenance and Troubleshooting

Inspecting the Cables and Cords

1
Examine all system cables, the power plug and cord for damage. Make sure that the prongs of the
plug do not move in the casing. If damaged, replace it with an appropriate Philips power cord.
Inspect the Measurement Link cable and ensure that it makes good connection with the MMS and
2
the FMS. Make sure that there are no breaks in the insulation.
Inspect the cable connecting the MMS to the monitor. Make sure the connectors are properly
3
engaged.
Inspect the patient cables, leads and their strain reliefs for general condition. Make sure there are
4
no breaks in the insulation. Make sure that the connectors are properly engaged at each end to
prevent rotation or other strain.
Apply the transducer or electrodes to the patient, and with the monitor switched on, flex the
5
patient cables near each end to make sure that there are no intermittent faults.

Maintenance Task and Test Schedule

The following tasks are for Philips-qualified service professionals only. All maintenance tasks and
performance tests are documented in detail in the service documentation supplied on the monitor
documentation DVD.
Ensure that these tasks are carried out as indicated by the monitor's maintenance schedule, or as
specified by local laws. Contact a Philips-qualified service provider if your monitor needs a safety or
performance test. Clean and disinfect equipment to decontaminate it before testing or maintaining it.
Maintenance and Test Schedule
Monitor Tests
Safety checks. Selected tests on the
basis of IEC 60601-1
Monitor Maintenance
Check ECG synchronization of the
monitor and defibrillator (only if
hospital protocol requires use of
monitor during defibrillation)
Replace backlight
(integrated displays only)
Multi-measurement Module and Module Tests
Performance assurance for all
measurements not listed below.
Multi-measurement Module and Module Maintenance
NBP calibration
Microstream CO2 calibration and
performance test
Mainstream and sidestream CO2
calibration check (M3014A)
366
Frequency
At least once every two years, or as needed, after any repairs
where the power supply is removed or replaced, or if the
monitor has been dropped.
At least once every two years, or as needed.
25,000 - 30,000 hours (about three years) of continuous
usage, or as needed.
At least once every two years, or if you suspect the
measurement values are incorrect.
At least once every two years, or as specified by local laws.
At least once a year or after 4000 operating hours.
At least once a year, or if you suspect the measurement
values are incorrect.

Advertisement

Table of Contents

Comments to this Manuals

Symbols: 0
Latest comments: