HP 8447D Operation And Service Manual page 28

Table of Contents

Advertisement

Figure 6
Operation and Service Manual
Performance Tests Procedures
HP 8447D Output VSWR, 1 to 110 MHz:
HP 8447E/F Output VSWR, 1 to 110 MHz:
HP 8447F Option H64 low frequency preamp
Output VSWR, 1 to 50 MHz:
VSWR Test Setup: 0.11 to 1.3 GHz
6 Connect the equipment as shown in Figure 6. Set the sweep oscillator for a leveled CW
signal, sweeping from 110 to 220 MHz. Connect the power meter to the s-parameter
test set INPUT PORT A and set the generator/sweeper output level for −15 dBm as read
on the power meter.
26
>8.8 dB
_____________
>7.7 dB
_____________
>9.6 dB
_____________