HP 8447D Operation And Service Manual page 24

Table of Contents

Advertisement

Operation and Service Manual
Performance Tests Procedures
Frequency
250 ±100 kHz (tune for maximum):
100 kHz:
For HP 8447F H64 low frequency preamp, replace the generator/sweeper shown in
Figure 4 with HP 3335A synthesizer (or equivalent). Set the synthesizer for a 9 kHz,
CW signal. Repeat step 7. Measure, calculate and record the amplifier gain at 9 kHz.
HP 8447F H64 Frequency
9 kHz:
9 Select the highest and lowest gains from steps 2, 4, 5, 7, and 8. Subtract the lowest gain
from the highest gain. The result is gain flatness and should be less than 3 dB.
Example:
23 dB (highest gain) minus 21.5 dB (lowest gain) = 1.5 dB (gain flatness)
For HP 8447F H64 low frequency preamp, select the highest and lowest gains from
steps 7, and 8. Subtract the lowest gain from the highest gain. The result is gain flat-
ness and should be less than 4 dB (±2 dB). Add the highest and lowest gains together
and divide by 2. The result is measured gain.
10 Add the highest and lowest gains together and divide by 2. The result is the measured
mean gain.
Example:
23 dB (highest gain)
+
21.5 dB (lowest gain)
----------------------------------------------------------------------------------------------------------- -
2
22
_______________dB
_______________dB
_______________dB
Gain flatness:__________________dB
Measured gain:__________________dB
=
22.25 dB (mean gain)
Gain
Gain