5-Wire Placement (EASI Mode)

5-20
ECG and Arrhythmia Monitoring
Make sure that the S and E electrodes line up vertically on
the sternum, and that the I, E and A electrodes align
horizontally.
1. E (V) on the lower sternum at the level of
the fifth intercostal space
2. A (LL) on the left midaxillary line at the
same level as the E electrode
3. S (LA) on the upper sternum
4. I (RA) on the right midaxillary line at the
same level as the E electrode
5. N (Reference) can be anywhere, usually
below the sixth rib on the right hip