ECG Leads Monitored

Depending on the patient cable connected to the MX40, a different set of
viewable leads are available at the MX40 and the Information Center. The
MX40 can source up to four raw ECG waves to the Information Center.
If you are using ...
3-wire
5-wire (Standard mode)
5-wire (EASI mode)
these leads can be selected at the MX40
and the Information Center
I, II, III
Sourced (raw) waves are received as:
Channel 1 = I, II, or III
Default is II.
I, II, III, aVR, aVL, aVF, MCL and V
Sourced (raw) waves are received as:
Channel 1 = II
Channel 2 = III
Channel 3 = MCL
Defaults are II, V, III.
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5,
V6
In EASI mode, the sourced (raw) waves
are received as:
Channel 1 = Vector 1 (A-I)
Channel 2 = Vector 2 (A-S)
Channel 3 = Vector 3 (E-S)
Defaults are II, V2, III, V5.
Arrhythmia monitoring is performed only on
the primary and secondary leads selected
at the Information Center, although you can
view and perform ST analysis on all 12
EASI derived leads.
ECG and Arrhythmia Monitoring
5-15