Alarm Text
TELE REMOVE BATT
Source - Monitor
TELE SERVICE BATT
Source - Monitor
TELE WEAK SIGNAL
Source - MX40
Priority
Condition
Hard,
The temperature of the
lithium-ion battery is
Latched
>60
o
C and the battery
must be removed.
Note —
For
Information Center
Release L or M, this
INOP will appear as
"REPLACE BATTERY
T".
Hard
The lithium-ion battery
has exceeded the
maximum charge cycle
limit.
Soft
Patient is at outer
range of the radio
coverage area.
The MX40 is
receiving a weak
signal with high data
loss from the AP.
Condition exists for
multiple devices in a
specific area
What to do
Replace the lithium-ion
battery.
Replace the lithium-ion
battery.
Return patient to the
coverage area.
If patient is in close
proximity to AP, replace
the MX40. Contact
service.
The AP covering the
specific area is suspect.
Contact Service.
Alarms
4-21