Philips IntelliVue MX40 Instructions For Use Manual Page 64

Wearable patient monitor.
Hide thumbs

Advertisement

Alarm Text
NO SIGNAL
(appears at the
Information Center only)
Source - Information
Center
REPLACE BATTERY T
Source - MX40
Note —
This INOP may
also be configured to
display as a Red or
Yellow Technical
Alarm.
SOME ECG ALARMS
OFF
Source - Information
Center
%SpO
T EQUIP MALF
2
Source - MX40
%SpO
T ERRATIC
2
Source - MX40
%SpO
T EXTD UPDATE
2
Numeric is replaced by a
-?-.
Source - MX40
4-18
Alarms
Priority
Condition
Hard,
The MX40 is outside
Latched
the coverage area,
or
No batteries in the
MX40, or
The MX40 has
failed.
Red or
Dead battery. No
Yellow or
monitoring is
Hard
occurring.
Technical
Alarm,
When operating
Latched
wirelessly, the
patient monitor is no
longer providing
power to the MX40,
and battery capacity
is now depleted.
Soft
Some yellow
arrhythmia alarms have
been turned off for this
patient.
Hard
Malfunction in the SpO
equipment
Hard
Erratic SpO
measurements, often
due to a faulty sensor
or invalid SpO
measurements, or
incorrect transducer
position
Soft
The update period of
displayed values is
extended due to an
NBP measurement on
the same limb or an
excessively noisy
signal.
What to do
Make sure that the MX40
is within the coverage
area and has good
batteries.
Replace the MX40 if
Power On Self Test fails.
Put bed in Standby.
Replace batteries.
For information only.
Contact Service.
2
Repeat measurement,
2
reposition sensor on patient,
or finally, replace sensor.
2
If NBP is not active, check
the sensor placement.
Reposition the sensor on
patient, or replace sensor.