Philips IntelliVue MX40 Instructions For Use Manual Page 63

Wearable patient monitor.
Hide thumbs

Advertisement

Alarm Text
LEADSET UNPLUGGED
Source - MX40
LOCAL AUDIO OFF
Source - MX40
MONIT. DISCONNECT
Source - MX40
NO ALARM DISPLAY
Source - MX40
NO CENTRAL
MONITOR
(appears at MX40 only)
Source - Patient
Monitor
Priority
Condition
Hard
Leadset has been
unplugged from the
MX40.
Soft
There is no alarm
audio notification when
operating in Telemetry
Mode.
Note —
This is normal
operation in Telemetry
Mode.
Hard
The paired
MX40/bedside monitor
is out of short-range
radio range or there is
excessive radio
interference.
Soft
When operating with
Information Center
Release L Or M, there
is no local alarming at
the MX40, networked
or non-networked.
Hard
The MX40 is out of
range of the
network.
Patient Sector at the
Information Center
is in Standby.
What to do
Re-attach the ECG leadset.
Change to Monitor Mode.
Reduce the distance
between the devices.
Identify and remove
interference source.
Condition is not present when
operating with Information
Center Release N or later
(unless specifically
configured to operate in this
way).
Return the MX40 to the
coverage area.
Select Resume at the
Information Center.
Alarms
4-17