Philips IntelliVue MX40 Instructions For Use Manual Page 62

Wearable patient monitor.
Hide thumbs

Advertisement

Alarm Text
cl NBP DISCONNECT
Source - Cableless
Measurement Device
cl SpO
Batt Low
2
Source - Cableless
Measurement Device
cl SpO
Batt Empty
2
Source - Cableless
Measurement Device
cl SpO
DISCONNECT
2
Source - Cableless
Measurement Device
ECG/ARRH ALARM
OFF
Source - MX40
ECG LEADS OFF
Note This INOP may
also be configured to
display as a Red or
Yellow Technical Alarm.
Source - MX40
<electrode> LEAD OFF
Source - MX40
4-16
Alarms
Priority
Condition
Hard
CL NBP Pod is not
connected with the
MX40.
Hard
CL SpO
Pod weak
2
battery condition.
Severe
CL SpO
Pod empty
2
battery condition.
Hard
CL SpO
Pod is not
2
connected with the
MX40.
Soft
ECG is turned off.
Red or
Multiple leads are
Yellow or
off.
Hard
Technical
Alarm
Hard
Single lead is off.
If primary lead is MCL,
lead will be identified
as V/C in INOP text.
What to do
Resolve interference
condition.
Reduce range between
CL NBP Pod and MX40.
Charge battery.
Charge battery. Monitoring is
not possible.
Resolve interference
condition.
Reduce range between
CL SpO
Pod and MX40.
2
Turn on ECG.
Re-attach ECG leads to
patient..
Re-attach ECG leads to
patient.