Philips IntelliVue MX40 Instructions For Use Manual Page 61

Wearable patient monitor.
Hide thumbs

Advertisement

Alarm Text
CANNOT ANALYZE
ECG
Source - MX40 and
Information Center
CHECK ECG SETTINGS
Source - MX40
CHECK PAIRING
Source - MX40
cl NBP Batt Low
Source - Cableless
Measurement Device
cl NBP Batt Empty
Source - Cableless
Measurement Device
Priority
Condition
Hard
Arrhythmia algorithm
cannot reliably analyze
the ECG data on any
monitored leads.
Hard
Synchronization of
ECG settings between
the monitor and the
Information Center has
failed.
Yellow
There is a problem
Technical
with device pairing.
Alarm
When the MX40 is
wirelessly paired
with an X2 patient
monitor (no label)
docked with a larger
networked MP
series monitor, and
the network
connection is lost.
Hard
CL NBP Pod weak
battery condition.
Severe
CL NBP Pod empty
battery condition.
What to do
Assess the lead selections,
initiate relearn, and validate
analyzed rhythm. Check
other INOPs for possible
source of problem.
Check that the ECG settings
from the ECG source are
appropriate.
Note —
When transitioning
between networked and
non-networked monitoring,
this INOP will display.
Pressing the Silence button
on the monitor will dismiss
the INOP.
Check that the bedside
monitor is correctly paired.
Select the correct device
to be paired.
Charge battery.
Charge battery. Monitoring is
not possible.
Alarms
4-15

Advertisement