Alarm Text
*** TACHY yyy > xxx
*** V-FIB/TACH
*** V-TACH
*/**AFIB
**NBP High
**NBP Low
** SpO
T yyy > xxx
2
** SpO
T yyy < xxxx
2
* HR yyy > xxx
* HR yyy < xxx
* IRREGULAR HR
* MISSED BEAT
Priority
Condition
Red
Extreme Tachycardia.
Heart Rate (yyy) greater than Extreme
Tachy limit
Red
Ventricular Fibrillation.
Fibrillatory wave (sinusoidal wave
between 2-10 Hz) for 4 consecutive
seconds
Red
Ventricular Tachycardia.
Consecutive PVCs greater than or
equal to V-Tach Run limit and Heart
Rate greater than V-Tach limit (xxx)
Yellow
Atrial fibrillation waveform detected
Yellow
High limit has been exceeded for high
pressure limit
Yellow
Low limit has been exceeded for low
pressure limit
Yellow
High SpO
.
2
SpO
value (yyy) greater than high
2
SpO
limit (xxx).
2
Yellow
Low SpO
.
2
SpO
value (yyy) less than low SpO
2
limit (xxx).
Yellow
Heart Rate (yyy) greater than the
upper Heart rate limit (xxx).
Yellow
Heart Rate (yyy) lower than the lower
Heart Rate limit (xxx).
Yellow
Consistently irregular rhythm (irregular
R-R intervals).
Yellow
No beat detected for 1.75 x average
R-R interval for Heart Rate greater
than 120, or no beat for 1 second with
Heart Rate greater than 120
(non-paced patient only).
Source
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
NBP
NBP
SpO
2
SpO
2
2
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
Alarms
4-11