AA Battery Charge Status
3-18
Basic Operation
Approximate
Approximate
Battery Life
Time
Remaining
Remaining
(ECG only)
Low battery
< 30 minutes
level to
replace/charg
e battery level
Replace/char
< 10 minutes
ge battery
level
Approximate
Approximate
Battery Life
Time
Remaining
Remaining
(ECG only)
100%
> 24 hours
75%
< 18 hours
50%
< 12 hours
25%
< 6 hours
10%
< 2 hours
Low battery
< 30 minutes < 30 minutes
level to
replace/charge
battery level
Replace/charg
< 10 minutes < 10 minutes
e battery level
Approximate
Functionality
Time
Disabled
Remaining
(ECG & Spo
2
Continuous)
< 30 minutes
SpO
and
2
short-range
radio are
disabled.
Display is at
half
brightness
< 10 minutes
Device
shutdown
Approximate
Functionality
Time
Disabled
Remaining
(ECG & Spo
2
Continuous)
> 9 hours
None
< 7 hours
None
< 5 hours
None
< 2 hours
None
< 1 hours
None
SpO
and
2
short-range
radio are
disabled.
Display is at
half
brightness.
Device
shutdown
Battery
Indicator
LCD
Segments
1 Red
Red Battery
Icon
Audio
1 Red
Red Battery
Icon
Battery
Indicator
LCD
Segments
5 Green
4 Green
3 Green
2 Green
1 Green
1 Red
Red Battery
Icon
Audio
1 Red
Red Battery
Icon