Lithium-ion Rechargeable Battery Care

Care of the rechargeable battery begins when you receive a new battery for
use and continues throughout the life of the battery. The table below lists
battery care activities and when they should be performed.
Activity
Perform a visual inspection.
Charge the battery.
Clean the battery
Charge stored batteries to at least
40% of their capacity every six
months.
Decommission the battery
Rechargeable batteries are charged using the IntelliVue CL Charging
Station. For information on charging station use, see Charging Li-ion
Rechargeable Batteries p. 12-7 .
Lithium-ion Rechargeable Battery Handling Precautions
Lithium-ion batteries store a large amount of energy in a small package.
Use caution when handling the batteries; misuse or abuse could cause
bodily injury and/or equipment damage.
Do not short circuit - take care that the terminals do not contact metal
(e.g. coins) or other conductive materials during transport and storage.
Do not crush, drop or puncture - mechanical abuse can lead to internal
damage and internal short circuits that may not be visible externally.
Do not apply reverse polarity.
Do not incinerate batteries or expose them to temperatures above 60
o
(140
F).
When to Perform
Before inserting a battery in the
MX40.
Upon receipt, after use, or if a low
battery state is indicated. To optimize
performance, a fully (or almost fully)
discharged battery should be charged
as soon as possible.
At each patient discharge, or in cases
when the battery is exposed to
contaminants.
When not in use for an extended
period of time.
When any of the following INOPs are
displayed on the MX40:
TELE SERVICE BATTERY
TELE CHECK BATT TEMP
TELE REMOVE BATT
Basic Operation
o
C
3-13