Fast Spo - Philips IntelliVue MX40 Instructions For Use Manual

Wearable patient monitor.
Hide thumbs

Advertisement

Characteristic
Channel Width
Trace Width
Aspect Ratio
Input Signal Reproduction Accuracy:
Overall Error
Frequency Response: Sinusoidal
Frequency Response: Triangular
Impulse Response:
(for waves marked with ST bandwidth)
Pacemaker Pulse Display Capability

FAST SpO

13-24 Safety Standards & Specifications
Performance Disclosure/Specification (in italics)
40 mm. Meets the ANSI/AAMI EC13 Section 3.2.9.7(a)
standard: minimum = 30mm.
0.3 mm. Meets the ANSI/AAMI EC13 Section 3.2.9.7(b)
standard: maximum = 1.0mm.
0.4s/mV. Meets the ANSI/AAMI EC13 Section 3.2.9.7(f)
standard: 0.4s/mV.
-2.9%. Meets the ANSI/AAMI EC13 Section 3.2.9.8(a)
standard: maximum = +/- 20%.
0.67 to 40 Hz (3 db down). Meets the ANSI/AAMI EC13
Section 3.2.9.8(b) standard: 0.67 to 40 Hz
(3 db down).
0 to 25% reduction. Meets the ANSI/AAMI EC13
Section 3.2.9.8(b) standard: 0 to 25% reduction.
Displacement = 0.08 mV, slope = 0.11 mV/s. Meets the
ANSI/AAMI EC13 Section 3.2.9.8(c) standard:
displacement maximum = 0.1 mV; slope
maximum = 0.30 mV/s.
Minimum = 0.2 mV RTI. Meets the ANSI/AAMI EC13
Section 3.2.9. 12 standard: minimum = 0.2 mV RTI.
2
Parameter
SpO
Measurement
2
Range (Calibration
and Display)
SpO
Accuracy
2
SpO
Resolution
2
Specification
0 to 100%
See table following.
1%