ECG Performance Disclosure/Specifications

Characteristic
Heart Rate Averaging Method
13-22 Safety Standards & Specifications
Parameter
Pacer Rejection
Performance
(Pace pulses with no
tails).
EMC Performance
Limits, radiated
immunity
ECG Patient Cable
Disconnection Safety
Performance Disclosure/Specification (in italics)
Two different methods are used:
Normally, heart rate is computed by averaging the 12
most recent RR intervals.
If each of 3 consecutive RR intervals are greater
than 1200 milliseconds (i.e. rate less than 50 b/min)
for adult and pediatric patients, then the 4 most
recent RR intervals are averaged to compute the HR.
Specification
Positive pacers
1
Amplitude
Width
+2 to +700 mV
0.1, 0.2, 0.5 and 1.0 ms
+2 to +500 mV
1.5 ms
+2 to +400 mV
2 ms
Negative pacers
1
Amplitude
Width
-2 to -700 mV
0.1, 0.2, 0.5 and 1.0
ms
-2 to -500 mV
1.5 ms
-2 to -400 mV
2 ms
1
Philips does not claim, verify, or validate
support for all available pacemakers.
Meets Essential Performance.
All ECG connections are patient safe within
750 msec of patient cable removal, with
patient leakage current <10
Exception:Leadset detection pins are
protected mechanically to prevent patient
contact.