MX40 Short-Range Radio

Parameter
Frequency Ranges
Radio Channel
assignment
Frequency Control
RF Output Power
MX40 Frequency
Accuracy during normal
operation
Modulation Type
Modulation Bandwidth
Specification
ISM Band: 2400-2483.5MHz
16 Radio Channel assigned, Fc= 2405
+5*(k-11)MHz, k=11,12,...,26
Configured via the bedside monitor.
-
1.5 to
-
4.5 dBm +0/-3dB (0.7 mW to 0.3 mW),
into Antenna load.
<+/-40ppm, includes temperature
compensation & aging effects
Direct Sequence Spread Spectrum (DSSS),
O-QPSK with half sine pulse shaping
modulation (1M40Q7D)
>500KHz, typically +/-950KHz (6dB Bandwidth),
typically +/-1.4MHz (20dB Bandwidth)
Safety Standards & Specifications
13-19