MX40 2.4 GHz Radio

2.4 GHz ISM
FCC and Industry Canada Radio Compliance: This device complies with
Part 15 of the FCC Rules and RSS-210 of Industry Canada. Operation is
subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation. Any
changes or modifications to this equipment not expressly approved by
Philips Medical Systems may cause harmful radio frequency interference
and void your authority to operate this equipment.
Parameter
Frequency Range
Channel Assignment
RF Output Power
Transceiver Frequency
Accuracy during normal
operation
Modulation Type
Modulation Bandwidth
Out of Band Spurious
Emission Levels
Specification
ISM Band: 2400 - 2483.5 MHz
48 radio channels assigned from 2401.056
MHz - 2482.272 MHz
Channel Spacing: 1.728 MHz
FCC: Channels 0-46 -17 dBm +/- 1 dB (40
mW to 63 mW, nominal 50 mW), into
antenna load. Channel 47 only - 14.5 dBm
+/- 1 dB.
ETSI: 12 dBm +/- 1 dB (13 mW to 20 mW,
nominal 16 mW), into antenna load
ARIB: 13.5 dBm +/- 1 dB (18 mW to 28
mW, nominal 22 mW), into antenna load
<+ 60 /- 100 KHz relative to channel
frequency, includes temperature
compensation and aging effects
GFSK, Gaussian Frequency Shift keying
(1M40Q7D)
Typically 1.4 MHz (20 dB Bandwidth)
Typically 980 KHz (6 dB Bandwidth)
Meets ETSI, RS210, FCC, ARIB standards