MX40 1.4 GHz Radio

1.4GHz WMTS (US only)
This device complies with Part 15 of the FCC Rules. Operation is subject to
the condition that this device does not cause harmful interference.
Operation of this equipment requires the prior coordination with a
frequency coordinator designated by the FCC for the Wireless Medical
Telemetry Service.
13-16 Safety Standards & Specifications
Parameter
Frequency Ranges
RF Output Power
(existing systems)
MX40 Frequency
Accuracy during normal
operation
Antenna Gain
Modulation Type
Out of Band Spurious
Emission Levels:
<= 1394 MHz, >= 1401
MHz
<= 1428 MHz, >= 1433
MHz
Occupied bandwidth as
defined by power in 99%
BW
Specification
Bands: 1395-1400 MHz and 1427-1432
MHz
Channel Spacing: 1.6 MHz
8 dBm +2/-1.5 dB (4.5 mW to 10 mW), into
antenna load
<+60/-100 KHz relative to channel
frequency, includes temperature
compensation and aging effects
-3 dBi
GFSK with Root Raised Cosine filtering
(1M60Q7D)
<-41 dBm in 1 MHz bandwidth for FCC
limit
< +/- 800 KHz