Advertisement

Immunity Test
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8

Recommended Separation Distance

Immunity Test
Conducted RF
IEC 61000-4-6
IEC 60601-1-2 Test
Level
3 A/m
Warning
The MX40, equipped with a wireless network interface, intentionally
receives RF electromagnetic energy for the purpose of its operation.
Therefore, other equipment may cause interference, even if that other
equipment complies with CISPR emission requirements.
In the following table, P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in meters (m).
Portable and mobile RF communications equipment should be used no
closer to any part of the MX40, including cables, than the recommended
separation distance calculated from the equation appropriate for the
frequency of the transmitter.
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than the compliance level in
each frequency range.
Interference may occur in the vicinity of equipment marked with this
symbol:
IEC 60601-1-2 Test
Level
3 Vrms
150 kHz to 80 MHz
Compliance Level
3 A/m
Compliance Level
3 VRMS
Safety Standards & Specifications
Electromagnetic
Environment Guidance
Power frequency magnetic
fields should be a t levels
characteristic of a typical
location in a typical
commercial and/or hospital
environment
Electromagnetic
Environment Guidance
Recommended separation
distance:
d = 1.2√P
13-9