Advertisement

Electromagnetic Emissions

Electromagnetic Immunity

Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
13-8
Safety Standards & Specifications
Warning
Using accessories other than those specified may result in increased
electromagnetic emission or decreased electromagnetic immunity of the
monitoring equipment.
Emissions Test
Radio Frequency
(RF) emissions
RF emissions
CISPR 11
Harmonized
emisssions
Voltage
fluctuations/Flicker
emissions
IEC 61000-3-3
The MX40 is suitable for use in the specified electromagnetic environment.
The user must ensure that it is used in the appropriate environment as
described below.
IEC 60601-1-2 Test
Level
±6 kV contact
±8 kV air
Compliance
Avoiding Electromagnetic
Interference
Group 1
TheMX40 uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
Class B
The MX40 is suitable for use in all
establishments.
Not Applicable
Device is battery powered only
Not Applicable
Compliance Level
±6 kV contact
±8 kV air
Electromagnetic
Environment Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.