Battery Status on the Charging Station Display

Battery Lifetime Management

12-8
Maintenance
Status
no battery on charger slot
battery put on charger slot
battery not properly recognized,
error
battery recognized, battery
charging
battery recognized, battery full
(>90%)
The AC Power / Error LED is
green when the charging station is connected to AC power
cyan during startup or to indicate a general charging station error
Note — Wiping of battery contacts with an alcohol solution after cleaning is
recommended.
The IntelliVue CL Charging Station display provides a quick overview of
all the connected devices and their battery status. The screen is arranged in
the same layout as the charger slots.
The lifetime of a li-ion battery depends on the frequency and duration of
use. When properly cared for, the useful life is approximately 4 years or 500
complete charge-discharge cycles, whichever comes first. In addition,
experience indicates that the incidence of failure may increase with battery
service life due to the accumulated stresses of daily use. We therefore
strongly recommend that li-ion batteries be replaced after 2 years or 500
complete charge-discharge cycles.
The age of a li-ion battery begins at the date of manufacture. The date of
manufacture is listed on the side of the battery.
LED
off
flashing yellow
cyan
yellow
green