Unit Settings - NBP Setup
Control
Function
Set patient type used for NBP
Patient Type
alarm limits.
Set NBP alarm notification.
NBP Alarm
Systolic High Increment/decrement NBP
high alarm limit by 1.
Systolic Low Increment/decrement NBP
low alarm limit by 1.
Increment/decrement NBP
Diastolic
high alarm limit by 1.
High
Diastolic Low Increment/decrement NBP
low alarm limit by 1.
Increment/decrement NBP
Mean High
high alarm limit by 1.
Increment/decrement NBP
Mean Low
low alarm limit by 1.
Settings
Adult
Pediatric
Systolic or Diastolic
Systolic
Diastolic
Mean
Off
Limit Maximum is 260
Limit Minimum is 160 (Adult)
Limit Maximum is 260
Limit Minimum is 75 (Ped.)
Limit Maximum is 155
Limit Minimum is 28 (Adult)
Limit Maximum is 255
Limit Minimum is 28 (Ped.)
Limit Maximum is 260
Limit Minimum is 90 (Adult)
Limit Maximum is 260
Limit Minimum is 45 (Ped.)
Limit Maximum is 85
Limit Minimum is 28 (Adult)
Limit Maximum is 40
Limit Minimum is 28 (Ped.)
Limit Maximum is 260
Limit Minimum is 65 (Adult)
Limit Maximum is 260
Limit Minimum is 55 (Ped.)
Limit Maximum is 60
Limit Minimum is 28 (Adult)
Limit Maximum is 50
Limit Minimum is 28 (Ped.)
Monitoring with the MX40 at the Information Center
Factory
Default
Adult
Systolic or
Diastolic
160 Adult
120 Pediatric
90 Adult
70 Pediatric
90 Adult
70 Pediatric
50 Adult
40 Pediatric
110 Adult
90 Pediatric
60 Adult
50 Pediatric
9-17