Unit Settings - Telemetry Setup
Control
SpO
Mode
2
Suppress
SpO
Inops
2
with NBP
9-14
Monitoring with the MX40 at the Information Center
Function
Determine the MX40 SpO
measurement behavior.
Note —
Pulse Rate and
Pleth Wave are not
available in Spot Check
(Manual) mode.
Enable/disable the SpO
algorithm to detect NBP
running and suppress
sending technical alarms
from the MX40 for 60
seconds.
Warning
If you measure SpO
on a
2
limb that has an inflated
NBP cuff, a non-pulsatile
SpO
technical alarm can
2
occur. If the monitor is
configured to suppress
this alarm, there can be a
delay of up to 60 seconds
in indicating critical patient
status, such as sudden
pulse loss or hypoxia.
Settings
Spot Check (Manual)-
2
Provides manual
measurements so the clinician
can check as needed.
Measurement can be initiated
from the SmartKeys menu, the
SpO
Setup menu or by
2
selecting the Spot Check
SpO
icon in the Patient
2
Window.
Continuous - Sends an SpO
parameter value to the
Information Center every
second. If selected, Pulse
Rate and Pleth Wave may
also be sent.
enable
2
disable
Factory
Default
Spot Check
(Manual)
2
enable