Patient-Configurable Settings in Telemetry Setup
Control
Function
Enable/disable the
Suppress
SpO
algorithm to
SpO
INOPs
2
2
suppress sending
with NBP
technical alarms from
the MX40 during an
NBP measurement for
60 seconds.
Warning
If you measure SpO
on a limb that has an
inflated NBP cuff, a
non-pulsatile SpO
technical alarm can
occur. If the monitor is
configured to suppress
this alarm, there can be
a delay of up to 60
seconds in indicating
critical patient status,
such as sudden pulse
loss or hypoxia.
Enable/disable the
Pleth Wave
transmission of the
Pleth wave (and its
subsequent display) to
the Information Center.
For Continuous SpO
mode only.
Enable/disable display
Pulse
of the Pulse rate at the
Information Center. For
Continuous SpO
only.
Turn SpO
SpO
Alarm
2
on/off at the
Information Center and
the MX40.
Change current
Unit Settings
settings back to last
saved clinical unit
settings.
Setting Choices
enable
disable
2
2
enable
disable
Note —
wave replaces the Vb wave in the
Patient Window during 6-lead
2
monitoring.
enable
disable
mode
2
alarms
enable (on)
2
disable (off)
(none)
Monitoring with the MX40 at the Information Center
When enabled, the Pleth
Factory
Default
enable
disable
(Pleth is not
displayed.)
disable
(Pulse rate
is not
displayed.)
enable
9-9