Physiological SpO
alarm signals will be generated. For adult patients, the
2
SpO
low limit can be set between 50 and 99% inclusive, in 1% increments,
2
and the SpO
high alarm limit can be set between 51 and 100% inclusive, in
2
1% increments. For pediatric patients, the SpO
low limit can be set between
2
30 and 99% inclusive, in 1% increments, and the high alarm limit can be set
between 31 and 100% inclusive, in 1% increments. The maximum delay
between the physiological alarm condition and alarm signal generation is
25 seconds.
Pulse rate is also derived from the pulsatile SpO
measurement, and
2
displayed results can range from 30 to 300 bpm. There is no alarm function
for pulse rate.
The pleth wave is auto-scaled to maximum display size. It decreases only
when the signal quality becomes marginal. Pleth wave size is NOT directly
proportional to the pulse volume.
7-4
SpO2 Monitoring