3. Depending upon the circumstance, replace either just the
particulate filter or both. Refer to
Moisture
Filters.
4. Reattach the pump inlet and replace the screw. Tighten
the screw using 3-4 in-lbs. torque. Do not overtighten.
5. Activate the detector to begin the startup test and the
pump block test. Depending upon the result, one of the
following screens display.
Successful: If the block test is successful, the detector
completes the startup test and enters normal operation.
Unsuccessful: If the block test is unsuccessful, the
detector deactivates. For causes and possible solutions,
refer to Troubleshooting.

Replacing the Pump

To obtain a new pump (XT-RPUMP-K1) and the Pump Replacement
Operator's Manual, refer to
Replacement Parts and
60
Particulate Filters
and
Accessories.