INOP Message, Indication
cl SpO₂ Batt Low
INOP tone
cl SpO₂ Batt Malf
INOP tone
cl SpO₂ Batt Temp
INOP tone
cl SpO₂ Check Batt
INOP tone
cl SpO₂ Chk ChrgIF
cl SpO₂ Chk Sett
cl SpO₂ Disconnect
INOP tone
cl SpO₂ No Cradle
INOP tone
cl SpO₂ Remove
INOP tone
cl SpO₂ Serv Batt
INOP tone
clResp License Req
clSpO₂ License Req
cmResp NeoPatient?
cmResp No Signal
cmRespCan'tAnalyze
cmRespPediPatient?
What to do
The remaining battery time of the SpO
There is a malfunction in the SpO
The temperature of the battery in the SpO
exposed to a heat source. If the INOP persists, remove the Pod from the patient and contact your
service personnel.
The battery in the SpO
Pod is nearing the end of its useful life. Only 50 charge/discharge cycles
2
remain. Contact your service personnel to replace the battery.
The interface between the SpO
the contacts and reposition the Pod on the charging station.
The SpO
Pod's settings may have changed or become corrupt. Check the settings.
2
The SpO
Pod has lost the connection to the monitor.
2
The SpO
Pod is not in its cradle.
2
The temperature of the battery in the SpO
and contact service personnel.
The battery in the SpO
Pod has reached the end of its useful life. It can no longer be charged. Contact
2
your service personnel to replace the battery.
The Respiration Pod requires a valid software license for operating.
The SpO
Pod requires a valid software license for operating.
2
The Respiration Pod cannot be used with neonatal patients. Check the patient category.
No signal is being detected from the Respiration Pod. Check that the Pod is attached correctly with the
adhesive attachment and that it moves with the patient's breathing.
The Respiration Pod algorithm cannot reliably analyze the signal. Respiration rate and pulse rate are
affected
The Respiration Pod cannot be used with pediatric patients. Check the patient category.
Pod is below 2 hours.
2
Pod's battery system. Contact your service personnel.
2
Pod is critically high. Check that the Pod is not covered or
2
Pod and the Charging Station is not working properly. Try cleaning
2
Pod is too high. Remove the SpO
2
4 Patient Alarms and INOPs
Pod from the patient
2
83