4 Patient Alarms and INOPs
C.O. INOPs
INOP Message, Indication
C.O. Deactivated
INOP tone
C.O. Equip Malf
Numeric is replaced by -?-
INOP tone
C.O. Unplugged
Numeric is replaced by -?-
INOP tone
CCI No BSA
CCI numeric unavailable
INOP tone
CCO BadPressSignal
Numeric is replaced by -?-
INOP tone
CCO No Calibration
Numeric is replaced by -?-
CCO No <Press Label>
Numeric is replaced by -?-
INOP tone may sound
CCO No Press
at Information Center
CCO Not Supported
Numeric is replaced by -?-
INOP tone
CCO Overrange
CCI Overrange
Numeric is replaced by -?-
INOP tone
CCO Press Invalid
at Information Center
CCO <Press Label> Invalid
Numeric is replaced by -?-
INOP tone may sound
CCO PressOverrange
Numeric is replaced by -?-
INOP tone
CCO PulseOverrange
Numeric is replaced by -?-
INOP tone
CCO Recalibrate
Numeric is replaced by -?-
80
What to do
The Cardiac Output measurement label in the measurement device has been deactivated by
Measurement Selection
deactivating the label in the
disappears from the display. To switch the measurement on again, reactivate the measurement label in
Measurement Selection
the
There is a problem with the C.O. hardware. Contact your service personnel.
Plug in the C.O. module. Silencing this INOP switches off the measurement.
CCI cannot be calculated because the patient's body surface area is unknown. Enter the patient weight
and height to provide the BSA for CCI calculation.
The arterial pressure wave can currently not be used for pulse contour calculation for CCO or CCI
measurement. Possible causes are air bubbles in the tubing or a physiological condition, for example
severe arrhythmia.
The CCO measurement is currently not calibrated.
CCO/CCI cannot be calculated. Make sure that the pressure chosen in the
CCO from
matches the pressure measured with the arterial catheter for CCO measurement. A
pressure from an external device cannot be used. Select another pressure label, either ABP, Ao, ART,
BAP, FAP, or UAP.
CCO/CCI cannot be calculated. Make sure that the pressure chosen in the
CCO from
matches the pressure measured with the arterial catheter for CCO measurement. A
pressure from an external device cannot be used. Select another pressure label, either ABP, Ao, ART,
BAP, FAP, or UAP.
A catheter for transpulmonary C.O. measurements has been unplugged and replaced with a Right
Heart C.O. catheter, or the measurement mode has been changed manually. Silencing this INOP
switches the measurement off.
The measured CCO or CCI value is not within the specified range for CCO/CCI measurement.
The arterial pressure selected for pulse contour calculation for CCO is available but currently invalid.
Make sure the pressure transducer is connected and the zero calibration is valid.
The arterial pressure selected for pulse contour calculation for CCO is available but currently invalid.
Make sure the pressure transducer is connected and the zero calibration is valid.
The mean value of the arterial pressure values used for pulse contour calculation for CCO is below
0 mmHg or above 300 mmHg.
The pulse rate of the pressure used for pulse contour calculation for CCO is below 30 bpm or above
240 bpm.
The most recent CCO or CCI calibration was made over 8 hours ago or the arterial pressure
measurement used for CCO calculation has been zeroed after the CCO calibration was performed.
You should recalibrate CCO or CCI with transpulmonary C.O. measurements at least every 8 hours or
when the hemodynamic condition of the patient has changed. The pressure measurement must be
zeroed before a CCO calibration.
window. The measurement automatically
window.
Setup CCO
menu under
Setup CCO
menu under