INOP Message, Indication
CO₂ No Sensor
from M3014A
Numeric is replaced by -?-
INOP tone
CO₂ No Tubing
Numeric is replaced by -?-
INOP tone
!! CO₂ Occlusion
!!! CO₂ Occlusion
Numeric is replaced by -?-
INOP tone
CO₂ Overrange
Numeric is replaced by -?-
INOP tone
CO₂ Pump Off
Numeric is replaced by -?-.
CO₂ Purging
Numeric is replaced by -?-
INOP tone
CO₂ Sens Warmup
Numeric is replaced by -?-
Microstream CO
: INOP tone
2
Mainstream CO
: no INOP tone
2
CO₂ Upgrade FW
Numeric is replaced by -?-
INOP tone
CO₂ Wait Cal2
Numeric is replaced by -?-
CO₂ Zero Failed
Numeric is replaced by -?-
INOP tone
CO₂Change Scale
CO₂NoTransducer
from mainstream CO
(except
2
M3014A)
Numeric is replaced by -?-
INOP tone
CO₂Zero Running
CO₂ZeroRequired
Numeric is replaced by -?-
INOP tone
IPI Check Pat. Age
IPI Check Sources
IPI Neo Patient ?
What to do
There is no CO
sensor connected. If you silence this INOP the CO
2
off.
Either the sample line is disconnected, or an incorrect line is attached. Check the connection. If
necessary, connect another sample line (Use only the approved accessories).
If you silence this INOP, the measurement will be switched off.
The sample line or exhaust tube is blocked. Check the tubing, then disconnect and reconnect the
sample line. If the INOP persists, connect a new sample line.
The CO
value is higher than the measurement range. If you suspect a false high value, contact your
2
service personnel.
The pump has been switched off for fifteen minutes. To switch it on again, select
Setup CO₂
menu.
The Filterline is being purged to remove an occlusion in the line or airway adapter. If the occlusion is
removed, the INOP will disappear. If not, the INOP
Wait until the sensor reaches operating temperature and the INOP disappears.
The software in the Measurement Extension does not match the software in the MMS. Contact your
service personnel.
Calibration on the first calstick cell is complete. Place the transducer on the other calstick cell and start
the CAL2 calibration cycle.
An error occurred during the last zero calibration. Check the airway adapter and clean, if necessary.
Perform another zero calibration. If the INOP persists, contact your service personnel.
The CO
wave is clipped. Select a more appropriate wave scale to display the whole wave.
2
There is no CO
transducer connected. If you replace the transducer, the new transducer must be
2
calibrated. If you silence this INOP the CO
Wait until zero calibration is finished.
Perform zero calibration for the CO
For Adult patients: a date of birth has been entered that results in a calculated patient age of less than
12 years. Enter correct date of birth or, if patient is pediatric, correct patient category.
For pediatric patients: either no date of birth has been entered (it is required for calculation of IPI for
pediatric patients) or the calculated age is below 1 year or above 12 years. Check that correct date of
birth is entered.
The IPI numeric cannot be provided as one or more of the measurement sources required for IPI is
not available. Check availability of etCO
Check patient category. The IPI numeric is not available for neonatal patients.
4 Patient Alarms and INOPs
CO₂ Occlusion
measurement will be switched off.
2
sensor. If the INOP persists, contact your service personnel.
2
, awRR, SpO
, and pulse rate.
2
2
measurement will be switched
2
Pump On
in the
is displayed.
79