4 Patient Alarms and INOPs
INOP Message, Indication
<Press Label> Noisy Signal
Pulse numeric is replaced by -?-
INOP tone
<Press Label> Overrange
Numeric is replaced by -?-
INOP tone
<Press Label> Unplugged
INOP tone
<Press Label> Zero+Check
Cal
Numeric is replaced by -?-
RAP
INOPs
UAP
INOPs
UVP
INOPs
CO2 INOPs
INOP Message, Indication
CO₂ Auto Zero
Numeric is replaced by -?-
if the Autozero lasts >15 sec,
INOP tone sounds.
CO₂ Cal Failed
Numeric is replaced by -?-
INOP tone
CO₂ Cal Mode
CO
numeric displays current
2
CO
value for accuracy check
2
CO₂ Cal Running
Numeric is replaced by -?-
CO₂ Check Cal
Numeric is replaced by -?-
INOP tone
CO₂ Chk Adapter
Numeric is replaced by -?-
INOP tone
CO₂ Deactivated
INOP tone
CO₂ Equip Malf
Numeric is replaced by -?-
INOP tone
78
What to do
This INOP can only arise when a pressure is selected as the pulse source. It occurs when the pulse
detector finds a pulse rate above 350 bpm. This is usually caused by movement artifact or electrical
interference.
Make sure that the measurement has been properly prepared and zeroed, and that the transducer is
level with the heart. If this INOP persists, try another transducer.
Possible causes are a measured pressure outside the allowed pressure range, or a broken wire to the
transducer.
A Pressure measurement label has been deactivated, either by unplugging a module, or by deactivating
Measurement Selection
the label in the
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module or reactivate the measurement label in
Measurement Selection
the
Perform a zero and check the calibration of the transducer.
<Press Label>
See
INOPs
<Press Label>
See
INOPs
<Press Label>
See
INOPs
What to do
The automatic zero calibration is in progress. This typically takes 10 seconds. During this time the CO
values may not be updated, or they may be replaced by -?-. Wait until the zero calibration is complete
to resume monitoring.
Make sure that the Cal cell was changed between CAL1 and CAL2. Repeat the calibration. If the
INOP reappears, try another transducer. If the INOP persists, contact your service personnel.
Currently no calibration is running. Accuracy can be checked by placing the transducer on the two cells
of the calstick and starting calibration. To start monitoring, leave Cal. Mode.
Wait until calibration is finished.
The CO
value is outside the measurement range. Perform an accuracy check for both calstick cells
2
and, if necessary, recalibrate the transducer.
Check that the sensor is connected to the airway adapter, clean the airway adapter, if necessary.
Perform a zero calibration. If the INOP persists, contact your service personnel.
The CO
measurement label in the measurement device has been deactivated by deactivating the label
2
Measurement Selection
in the
To switch the measurement on again, reactivate the measurement label in the
Selection
window.
The Measurement Extension is faulty. Unplug and replug the Multi-Measurement Module with
Extension. If you are using the mainstream method, unplug and replug the transducer or try another
transducer. If the INOP persists, contact your service personnel.
window.
window.
window. The measurement automatically disappears from the display.
2
Measurement