4 Patient Alarms and INOPs
INOP Message, Indication
<SpO₂ Label> Erratic
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Extd.Update
Numeric is displayed with a ?
(questionable numeric)
<SpO₂ Label>Interference
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Low Perf
Numeric is displayed with a ?
(questionable numeric)
<SpO₂ Label> No Pulse
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> No Sensor
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> NoisySignal
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Poor Signal
Numeric is displayed with a ?
(questionable numeric)
<SpO₂ Label> Pulse?
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Searching
Numeric is unavailable
<SpO₂ Label> Sensor Malf
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Sensor Off
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Unkn.Sensor
Numeric is replaced by -?-
<SpO₂ Label> Unplugged
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Upgrade
Numeric is replaced by -?-
76
What to do
Check the sensor placement. Try another adapter cable and sensor. If the INOP persists, contact your
service personnel.
The update period of displayed values is extended due to an NBP measurement on the same limb or
an excessively noisy signal.
There is too much interference, caused by a high level of ambient light and/or electrical interference.
Cover the sensor to minimize ambient light. If the INOP persists, make sure that the sensor cable is
not damaged or positioned too close to power cables.
Accuracy may be compromised due to very low perfusion. Stimulate circulation at sensor site. If INOP
persists, change the measurement site.
Check the perfusion at measurement site. If necessary, stimulate circulation or change measurement
site. If the INOP is due to NBP measurement on the same limb, wait until the NBP measurement is
finished.
Make sure the SpO
sensor is connected. If the INOP persists, try another adapter cable and sensor. If
2
you silence this INOP, the measurement will be switched off.
Excessive patient movement or electrical interference is causing irregular pulse patterns. Try to reduce
patient movement or to relieve the cable strain on the sensor.
The signal condition of the SpO
compromised. If the INOP persists, consider changing the application site or using another sensor.
The detectable pulsations of the SpO
SpO
is analyzing the patient signal to derive Pulse, SpO
2
analysis is complete.
The SpO
sensor or adapter cable is faulty. Try another adapter cable and sensor. If the INOP persists,
2
contact your service personnel.
The SpO
sensor is not properly applied to the patient. Apply the sensor following the instructions
2
supplied by the manufacturer.
The connected sensor or adapter cable is not supported by the SpO
sensors and cables.
An SpO
measurement label has been deactivated, either by unplugging a module, or by deactivating
2
Measurement Selection
the label in the
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module or reactivate the measurement label in
Measurement Selection
the
The SpO
measurement is currently in upgrade mode. Monitoring is not possible in this mode.
2
measurement is poor and measurement accuracy may be
2
signal are outside the specified pulse rate range.
2
and Perf values. Please wait until the search
2
window.
window.
measurement. Use only specified
2